Cloud Defense Logo

Products

Solutions

Company

Book A Live Demo

CVE-2017-12726 Explained : Impact and Mitigation

Discover the impact of CVE-2017-12726, a hard-coded password vulnerability in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump. Learn about affected versions and mitigation steps.

An issue with hard-coded passwords has been found in the Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, specifically affecting Versions 1.1, 1.5, and 1.6. This vulnerability allows unauthorized access to the pump's Telnet feature, potentially compromising external communications.

Understanding CVE-2017-12726

This CVE identifies a Use of Hard-coded Password vulnerability in the Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump.

What is CVE-2017-12726?

The CVE-2017-12726 vulnerability involves hardcoded credentials in the Telnet feature of the Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, impacting Versions 1.1, 1.5, and 1.6.

The Impact of CVE-2017-12726

The vulnerability allows unauthorized users to access the pump's Telnet feature, potentially compromising external communications. However, Smiths Medical states that file uploads via Telnet are not possible, limiting the impact to the communications module.

Technical Details of CVE-2017-12726

This section provides detailed technical information about the vulnerability.

Vulnerability Description

The issue involves hardcoded passwords in the Telnet feature of the affected pump versions, enabling unauthorized access.

Affected Systems and Versions

        Product: Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump
        Versions: 1.1, 1.5, and 1.6

Exploitation Mechanism

Unauthorized users can exploit the hardcoded credentials in the Telnet feature to gain access to the pump, potentially compromising external communications.

Mitigation and Prevention

Protective measures to address the CVE-2017-12726 vulnerability.

Immediate Steps to Take

        Disable Telnet access if not required for operation
        Implement strong, unique passwords for all system access
        Regularly monitor and audit system access logs

Long-Term Security Practices

        Conduct regular security assessments and penetration testing
        Keep systems up to date with the latest security patches

Patching and Updates

        Apply patches and updates provided by Smiths Medical to address the hardcoded password vulnerability