Learn about CVE-2018-14781 affecting Medtronic insulin pumps, allowing unauthorized insulin delivery. Find mitigation steps and long-term security practices.
Medtronic insulin pump models are vulnerable to a capture-replay attack, allowing unauthorized insulin delivery.
Understanding CVE-2018-14781
What is CVE-2018-14781?
The Medtronic MMT 508 MiniMed insulin pump and other models are susceptible to a capture-replay attack, enabling attackers to initiate insulin delivery.
The Impact of CVE-2018-14781
The vulnerability in Medtronic insulin pumps can lead to unauthorized insulin administration, posing serious health risks to patients.
Technical Details of CVE-2018-14781
Vulnerability Description
The vulnerability allows attackers to capture wireless transmissions between the pump and remote controller, replaying them to trigger insulin delivery.
Affected Systems and Versions
Exploitation Mechanism
Mitigation and Prevention
Immediate Steps to Take