Learn about CVE-2019-10964 affecting Medtronic MinMed 508 and Paradigm Series Insulin Pumps. Understand the risks, affected versions, and mitigation steps.

The Medtronic MinMed 508 and Medtronic Minimed Paradigm Insulin Pumps are vulnerable to unauthorized data manipulation due to a lack of proper authentication in their wireless RF communication protocol.

Understanding CVE-2019-10964

What is CVE-2019-10964?

The CVE-2019-10964 vulnerability affects Medtronic insulin pumps, allowing attackers in close proximity to manipulate data and potentially take control of insulin delivery.

The Impact of CVE-2019-10964

The vulnerability enables attackers to inject, replay, modify, or intercept data, posing a serious risk of altering pump settings and controlling insulin administration.

Technical Details of CVE-2019-10964

Vulnerability Description

The affected Medtronic insulin pumps utilize a wireless RF communication protocol lacking proper authentication, making them susceptible to unauthorized data manipulation.

Affected Systems and Versions

        Medtronic MiniMed 508 pump All versions
        MiniMed Paradigm 511 pump All versions
        MiniMed Paradigm 512/712 pump All versions
        MiniMed Paradigm 712E pump All versions
        MiniMed Paradigm 515/715 pumps–All versions
        MiniMed Paradigm 522/722 pump–All versions
        MiniMed Paradigm 522K/722K pumps–All versions
        MiniMed Paradigm 523/723 pumps–Software versions 2.4A or lower
        MiniMed Paradigm 523K/723K pumps versions 2.4A or lower
        MiniMed Paradigm Veo 554/754 pumps–versions 2.6A or lower
        MiniMed Paradigm Veo 554CM and 754CM versions 2.7A or lower

Exploitation Mechanism

Attackers with physical proximity to the affected insulin pumps can exploit the vulnerability to manipulate data, potentially leading to unauthorized control over insulin delivery.

Mitigation and Prevention

Immediate Steps to Take

        Disable wireless RF communication on the affected insulin pumps if not essential for treatment.
        Follow Medtronic's guidance on secure pump operation and data protection.

Long-Term Security Practices

        Regularly update pump software and firmware to mitigate known vulnerabilities.
        Implement strong physical security measures to prevent unauthorized access to the insulin pumps.

Patching and Updates

        Stay informed about security advisories and patches released by Medtronic to address the vulnerability.