CVE-2020-25164 : Exploit Details and Defense Strategies

Discover the vulnerability in B. Braun Melsungen AG's SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus products allowing unauthorized access to administrative interface user credentials. Learn about the impact, affected systems, exploitation, and mitigation steps.

A vulnerability in B. Braun Melsungen AG's SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus products allows attackers to recover user credentials of the administrative interface.

Understanding CVE-2020-25164

This CVE involves vulnerabilities in various products by B. Braun Melsungen AG, potentially compromising user credentials.

What is CVE-2020-25164?

The vulnerability in SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus products enables attackers to retrieve administrative interface user credentials.

The Impact of CVE-2020-25164

The vulnerability poses a medium severity risk with high confidentiality impact, requiring user interaction for exploitation.

Technical Details of CVE-2020-25164

This section delves into the specifics of the vulnerability.

Vulnerability Description

The flaw in B. Braun Melsungen AG's SpaceCom Version L81/U61 and earlier, and Data module compactplus Versions A10 and A11 allows unauthorized access to user credentials.

Affected Systems and Versions

        SpaceCom: Versions L81/U61 and earlier
        Battery Pack with Wi-Fi: Versions L81/U61 and earlier
        Data module compactplus: Versions A10 and A11

Exploitation Mechanism

Attackers can exploit this vulnerability over a network with low complexity, requiring user interaction.

Mitigation and Prevention

Protective measures and solutions to address the CVE.

Immediate Steps to Take

        Apply updates: SpaceCom - Version U62 or later (US), L82 or later (outside US)
        Battery Pack SP with Wi-Fi - Version U62 or later (US), L82 or later (outside US)
        Data module compactplus - Version A12 or later

Long-Term Security Practices

        Restrict direct Internet access to devices
        Implement a firewall and isolate medical devices from the business network

Patching and Updates

Contact B. Braun for assistance and refer to the B. Braun Security Advisory for detailed information.